Atomic Absorption Spectrometer Atomic Absorption Spectrometer Atomic Spectroscopy 

Place of Origin: Zhejiang, China (Mainland) 
inquire
Add to My Favorites
HiSupplier Escrow
Share |

Product Detail

Model No.: AA500F Atomic Absorption Sp...


 
 
<span style="f

 

 

Food Flame Atomic Absorption Spectrometer AA500F, Wavelength Reproducibility <0.05nm

 

 

Flame Atomiser Features

 

The flame atomizer offers three flame options: Air/Acetylene is the standard configuration with the N2O/acetylene and Air/LPG as options.

 

 

Air/Acetylene

 

The Air/Acetylene flame uses a 100mm single slot burner for the standard configuration

The high sensitivity (Cu 2ppm>0.280Abs) is due to the efficiency of the fixed position glass nebulizer fitted as standard. An acid resistant replacement is available as an option.

The flame can be easily set from blue lean flame through stoichiometric to fuel rich by means of computer control

 

 

N2O/Acetylene

 

The N2O/Acetylene flame uses a 50mm single slot burner and is available as an optional extra.

This flame configuration is used to measure elements less prone to ionization such as Aluminum, Tin, Titanium, Calcium, Vanadium and Molybdenum.

Switching from Air/Acetylene to N2O/Acetylene to Flame Off is fully controlled by the AAWin Pro software

 

 

Air/Propane (LPG)

This flame uses a 3 slot burner and with the low pressure requirement it is also much safer to operate

Due to the low temperature of the flame it is ideal for analyzing alkali metals such as Potassium, Sodium and Lithium, especially when used in the emission mode.

Some remote areas of the world have difficulty obtaining Acetylene gas of a high enough purity to operate the flame correctly. LPG can give a real alternative and offer comparable results for most elements throughout the wavelength range.

 

 

Safety Features

 

Pressure monitoring for all gases

Burner identification

Flame sensor

Drain trap level sensor

Gas leak detector

Over pressure in premix

Safety cut off switch

 

 

Applications:

 

Agricultural

Clinical

Environmental

Food

Metal

Mining

Geological

Pertrochemical

Pharmaceutical

 

 

Specifications:

 

 

Spectrometer System

Wavelength range

190-900nm

Light source

Hollow cathode lamp (HCL), Deuterium Arc lamp (D2)

Modulation

Square wave pulse

Modulation frequency

100hz self reversal (SR) background, 400hz Deuterium Arc (D2) background

Monochromator

Grating

1800 grooves/mm diffraction grating

Blazed wavelength

250nm

Focus

300nm

Bandwidth

0.1, 0.2, 0.4, 1.0, 2.0nm

Scan mode

automatic

Photometric type

Single beam

Wavelength accuracy

+0.15nm

Wavelength resolution

0.2nm+0.02nm

Wavelength reproducibility

<0.05nm

Baseline stability

0.005A/30min

Background correction

Deuterium Arc (D2) 1.0Abs, Self Reversal (SR) 3.0Abs

Flame Analysis

Flame types

Air/Acetylene, N

bigPhoto
 Mr. David Wang
Sale Manager
Tel: 86-10-69910653
Fax: 86-10-69910516
Contact to this supplier

Didn't find what you're looking for? Post Buying Lead or contact HiSupplier Customer Service Center  for help!

Related Search

Find more related products in following catalogs on Hisupplier.com

Company Info

PG Instruments Limited [China (Mainland)]


Business Type:Manufacturer
City: Beijing
Province/State: Beijing
Country/Region: China (Mainland)

You May Like: